Student Work

Ultrasound-Guided Needle Insertion Device for PCNL

Public Deposited

Downloadable Content

open in viewer

This project is a continuation of a proposed system that aligns the needle with the ultrasound image to improve the accuracy of Percutaneous Nephrolithotomy (PCNL) procedure. The device consists of an ultrasound probe, probe holder, acoustic reflector attachment, and optimal materials. Through image processing and analysis, 3% agarose was selected as the filling and stainless steel was selected as the optimal reflector as a replacement for glass mirrors. The attachment design was redesigned to a vertical position to improve the handling of the device and components were designed for manufacturability. These design changes proved successful during mock phantom and needle visualization testing done on the final version of the device. Finally, a set of improvements was recommended for future groups working on this device.

  • This report represents the work of one or more WPI undergraduate students submitted to the faculty as evidence of completion of a degree requirement. WPI routinely publishes these reports on its website without editorial or peer review.
Creator
Subject
Publisher
Identifier
  • 121595
  • E-project-042424-204631
Advisor
Year
  • 2024
UN Sustainable Development Goals
Date created
  • 2024-04-24
Resource type
Major
Source
  • E-project-042424-204631
Rights statement

Relations

In Collection:

Items

Items

Permanent link to this page: https://digital.wpi.edu/show/nz806402t