Design of a Sternal Fixation Device and Bicortical Bone Model
PúblicoConteúdo disponível para baixar
open in viewerFailure to properly close the sternum after a sternotomy can lead to complications. Current sternal fixation devices, including screw and plate systems, reduce the risk of complications but are unable to provide a flush fit with the sternum and prevent screw loosening. The first goal of this study was to optimize a screw and plate system that addresses this issue. The system utilizes a lag-lock mechanism that permits locking of the screw after the plate is flush fit with the bone. The second goal was to design and characterize a sternal model to mimic the mechanical properties of bone. It was characterized using axial and lateral pullout tests and compared to a standard bone model. In testing, the custom model performed similar to how bone is expected to, making it a better sternal model.