Student Work
Active power-line interference attenuation in bioamplifiers
公开In this project, we have implemented a feedback circuit that reduces common mode electrical potentials on the human body, in order to improve the performance of bioelectric sensors. The technique, known as "driven right let" (DRL), uses a feedback-based interference cancellation method. The bioelectric sensors used were inexpensive, high impedance, electrode-amplifiers that were electrically isolated from the recorder. An interference attenuation of at least 10 times was observed. This report serves as a reference into DRL and bioamplifier circuits.
- This report represents the work of one or more WPI undergraduate students submitted to the faculty as evidence of completion of a degree requirement. WPI routinely publishes these reports on its website without editorial or peer review.
- Creator
- Publisher
- Identifier
- 02D008M
- Advisor
- Year
- 2002
- Date created
- 2002-01-01
- Resource type
- Major
- Rights statement
- 最新修改
- 2021-09-14
关系
- 属于 Collection:
项目