Student Work
External Coil for MR Imaging of the Prostate
Pubblico DepositedContenuto scaricabile
open in viewerThe goal of this project was to show that the Helmholtz coil architecture is viable for use in MRI systems. Based on DC simulations, we have determined the coil dimensions to accommodate the largest range of patients, fit in an MRI scanner, and have the best possible SNR. Through exhaustive bench testing, we verified that our coil was able to be tuned to human loads and exhibited expected behavior, supporting that this product should be pursued in the future.
- This report represents the work of one or more WPI undergraduate students submitted to the faculty as evidence of completion of a degree requirement. WPI routinely publishes these reports on its website without editorial or peer review.
- Creator
- Publisher
- Identifier
- E-project-030610-122917
- Advisor
- Year
- 2010
- Date created
- 2010-03-06
- Resource type
- Major
- Source
- E-project-030610-122917
- Rights statement
- Ultima modifica
- 2023-11-01
Relazioni
- In Collection:
Articoli
Elementi
Thumbnail | Titolo | Visibilità | Embargo Release Date | Azioni |
---|---|---|---|---|
External_Coil_for_MR_Imaging_of_the_Prostate.pdf | Pubblico | Scaricare |
Permanent link to this page: https://digital.wpi.edu/show/5999n788t