Designing Non-Pneumatic Anti-Shock Trousers
PublicDownloadable Content
open in viewerHemorrhagic shock is a significant cause of death for trauma patients. Current treatment methods face a variety of shortcomings and limitations. The team conducted research and testing which informed the design and prototyping of a device to combat hemorrhagic shock by shunting blood from the lower extremities to the core of the body. The team decided to create a fully functional segment of the device to study proof of concept and create a baseline for the future of the design. Verification testing showed that the device does not obstruct the cutaneous view. Furthermore, the team found the device can be applied within the identified time window of less than five minutes. It was measured that the device can apply 30-50 mmHg which is in the desired range of pressure to meet the device objectives.
- This report represents the work of one or more WPI undergraduate students submitted to the faculty as evidence of completion of a degree requirement. WPI routinely publishes these reports on its website without editorial or peer review.
- Creator
- Publisher
- Identifier
- E-project-050421-082044
- 21741
- Keyword
- Advisor
- Year
- 2021
- Date created
- 2021-05-04
- Resource type
- Major
- Rights statement
- Last modified
- 2021-08-28
Relations
- In Collection:
Items
Items
Thumbnail | Title | Visibility | Embargo Release Date | Actions |
---|---|---|---|---|
Designing Non-Pneumatic Anti-Shock Trousers MQP Report.pdf | Public | Download |
Permanent link to this page: https://digital.wpi.edu/show/3j3335285