Etd
Measuring the TG-43 Parameters of Iridium-192 using Monte Carlo-based Dosimetry
PúblicoContenido Descargable
open in viewerRadioactive sources used in brachytherapy must be dosimetrically characterized prior to clinical use as defined the TG-43 protocol. In our previous project, Gafchromic film dosimetry was used to experimentally obtain the anisotropy function for an M-19 iridium-192 brachytherapy seed being developed by Source Production Equipment Corp (St. Rose, LA). In this project, the Monte Carlo N-Particle Transport code (MCNP) was used to computationally obtain the full set of TG-43 parameters including the Dose Rate Constant, the Reference Dose Rate, the Radial Dose Function, and the Anisotropy Constant for the M-19 seed.
- Creator
- Colaboradores
- Degree
- Unit
- Publisher
- Identifier
- etd-2341
- Palabra Clave
- Advisor
- Orcid
- Committee
- Defense date
- Year
- 2019
- Date created
- 2019-08-03
- Resource type
- Rights statement
- Última modificación
- 2023-09-20
Las relaciones
- En Collection:
Elementos
Elementos
Miniatura | Título | Visibilidad | Embargo Release Date | Acciones |
---|---|---|---|---|
MS_Thesis_Paper-Final_1.pdf | Público | Descargar |
Permanent link to this page: https://digital.wpi.edu/show/dz010s228